Three-dimensional bioprinting using a coaxial needle with viscous inks in bone tissue engineering - An In vitro study
Java Walladbegi1, Christian Schaefer1, Elin Pernevik2, Sanna Sämfors2, Göran Kjeller1, Paul Gatenholm2, George K Sándor3, Lars Rasmusson4
1 Department of Oral and Maxillofacial Surgery, Institute of Odontology, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden 2 Wallenberg Wood Science Center, Department of Chemistry and Chemical Engineering, Chalmers University of Technology, Gothenburg, Sweden 3 Department of Oral and Maxillofacial Surgery, Medical Research Center, University of Oulu, Oulu University Hospital, Oulu, Finland 4 Department of Oral and Maxillofacial Surgery, Institute of Odontology, The Sahlgrenska Academy, University of Gothenburg, Gothenburg; Department of Oral and Maxillofacial Surgery, Linkoping University Hospital, Linkoping, Sweden
Date of Submission | 25-Jun-2020 |
Date of Acceptance | 09-Oct-2020 |
Date of Web Publication | 23-Dec-2020 |
Correspondence Address: Dr. Java Walladbegi Medicinaregatan 12D, PO Box 450, 405 30 Gothenburg Sweden
  | Check |
DOI: 10.4103/ams.ams_288_20
Introduction: Vascularized autologous tissue grafts are considered “gold standard” for the management of larger bony defects in the craniomaxillofacial area. This modality does however carry limitations, such as the absolute requirement for healthy donor tissues and recipient vessels. In addition, the significant morbidity of large bone graft is deterrent to fibula bone flap use. Therefore, less morbid strategies would be beneficial. The purpose of this study was to develop a printing method to manufacture scaffold structure with viable stem cells. Materials and Methods: In total, three different combinations of ground beta tri-calcium phosphate and CELLINK (bioinks) were printed with a nozzle to identify a suitable bioink for three-dimensional printing. Subsequently, a coaxial needle, with three different nozzle gauge combinations, was evaluated for printing of the bioinks. Scaffold structures (grids) were then printed alone and with additional adipose stem cells before being transferred into an active medium and incubated overnight. Following incubation, grid stability was evaluated by assessing the degree of maintained grid outline, and cell viability was determined using the live/dead cell assay. Results: Among the three evaluated combinations of bioinks, two resulted in good printability for bioprinting. Adequate printing was obtained with two out of the three nozzle gauge combinations tested. However, due to the smaller total opening, one combination revealed a better stability. Intact grids with maintained stability were obtained using Ink B23 and Ink B42, and approximately 80% of the printed stem cells were viable following 24 hours. Discussion: Using a coaxial needle enables printing of a stable scaffold with viable stem cells. Furthermore, cell viability is maintained after the bioprinting process. Keywords: Adipose stem cells, bioinks, CELLINK, in vitro, viability
How to cite this article: Walladbegi J, Schaefer C, Pernevik E, Sämfors S, Kjeller G, Gatenholm P, Sándor GK, Rasmusson L. Three-dimensional bioprinting using a coaxial needle with viscous inks in bone tissue engineering - An In vitro study. Ann Maxillofac Surg 2020;10:370-6 |
How to cite this URL: Walladbegi J, Schaefer C, Pernevik E, Sämfors S, Kjeller G, Gatenholm P, Sándor GK, Rasmusson L. Three-dimensional bioprinting using a coaxial needle with viscous inks in bone tissue engineering - An In vitro study. Ann Maxillofac Surg [serial online] 2020 [cited 2021 Jan 27];10:370-6. Available from: https://www.amsjournal.com/text.asp?2020/10/2/370/304405 |
Introduction | |  |
Many strategies have been proposed for the management of craniomaxillofacial congenital bony defects or those defects following avulsive trauma or ablative surgery;[1] nevertheless, autogenous bone grafts are still considered “gold standard” for reconstructive bone surgery.[2]
Small-sized mandibular defects ranging from less than 4 to 6 cm can be managed with nonvascularized corticocancellous grafts harvested from different sites, for example, the anterior or posterior iliac crest.[2],[3],[4] Larger defects are commonly treated with vascularized grafts such as the fibula,[2],[5],[6] which offer a long pedicle, wide vessel diameter, and the possibility to incorporate skin, muscle, and bicortical bone components if such tissues are required.[7] However, the use of vascularized autologous tissue grafts requires healthy donor tissues and vessels, as well as recipient vessels. The significant morbidity of large bone graft harvesting is deterrent to fibula bone flap use.[8],[9],[10],[11] Less morbid treatment strategies would be advantageous,[12] although, in a longer perspective, the quality of life among these patients seems to be acceptable.[13]
Bone tissue engineering aims to regenerate osseous tissue by combining biomaterials and stem cells.[12],[13],[14],[15],[16],[17] One reliable source of stem cells for bone regeneration is autologous fat tissue that provides an abundant amount of adipose stem cells (ASCs).[17],[18] ASCs are capable of multiple lineage differentiation to adipocyte chondrocyte and osteoblast pathways.[17] Harvesting fat tissue is simple and causes far less morbidity in comparison to traditional bone harvesting techniques.[17],[18],[19],[20],[21] Another advantage of ASCs compared to bone marrow stromal/stem cells (BMSCs) is that BMSCs are present in low frequency in the bone marrow, whereas ASCs can be retrieved in high numbers from adipose tissue and can easily be expanded in vitro.[16],[22] Using the tissue engineering model,[14] it has been possible to harvest autogenous ASCs from patients and to use the cells to seed a resorbable scaffold.[23],[24],[25],[26]
Biomaterials of various compositions and physical forms can serve as scaffolds in bone tissue engineering, including granular forms and solid blocks. Although they have no load bearing ability compared to solid blocks, granules present more surface area for cellularization. In addition, granules are more quickly incorporated into host tissues than block forms, which remodel more slowly.[27] Granular calcium phosphate ceramics and related biomaterials such as hydroxyapatite possess macro- and micro-structural properties that make them useful biomaterials in tissue engineering. Such properties affect cell attachment, survival, signaling, growth, and propagation.[27],[28] While some inorganic ceramics may also bind directly to bone,[29] the surfaces of other biomaterials may affect differentiation of stem cells to osteoblasts.[30]
One obstacle in planning complex reconstructive surgery using either autologous bone or tissue-engineered constructs is the inability to produce complex facial contours in a predictable manner, using commercially available reconstruction plates and meshes.[31] One solution to this problem is to use computer-guided surgical planning and additive manufacturing technology to produce a passive, but fitting implant, designed for patient-specific anatomical needs.[32],[33] Yet, complex defects are difficult to restore with existing three-dimensional (3D) technology. The development of simultaneous 3D printing of scaffolds with cells for medical use is one technique that shows great promise.
A variety of 3D printing processes are currently being used. These include, inkjet, orifice free, and extrusion bioprinting.[33] Many of these approaches allow for the printing of supportive scaffold structure, as well as the simultaneous printing of different types of cells. However, there are several problems related to the use and the composition of the specific bioink, printing technology, and the type of cell laden structure to be produced. These include the primary stability, the dispersion of cells, and the specific combination of different biomaterials used in producing a viable implant. Another well-known problem is the viscosity versus clogging and pressure effects on the printed cell populations together with mechanical stability of the printed construct.[34] Combining stem cells, growth, and linking factors together with a stable scaffold material to replace complex-shaped bony resection defects is the primary goal of bioprinting.
Different combinations of materials have been used in additive manufacturing from cellulose to tri-calcium phosphate (TCP) and hydrogel-based compounds.[34] The addition of a sacrificial or resorbable material used initially for extra stability has been advocated, and cross-linking hydrogel is the most commonly used technique. In large defects, a more rigid and stable material such as TCP would be of benefit.
The aim of the present study was to develop a printing method to achieve a stable scaffold structure with viable stem cells.
Materials and Methods | |  |
The current in vitro study was performed within August 2018 to October 2019. In total, four independent experiments were carried out to address the aim of the present study. An initial experiment compared bioinks regarding printability; the second experiment investigated a coaxial needle for printing resolution, using different nozzle gauge configurations, and the third experiment measured the stability of printed grids [Figure 1]. In the fourth experiment, cell viability of the printed grids comprising human adipose tissue-derived mesenchymal stem cells (hAD-MSCs) was evaluated. In addition, cell viability of the same grids but at two different pressures was assessed.
Inks and bioinks
To identify suitable bioinks for extrusion printing, three different combinations (Ink B23; Ink B42, and Ink B48) of ground β-TCP (ChronOS Bone Graft, Synthes, West Chester, PA, USA) and nanofibrillated cellulose/alginate (NFC-A; CELLINK AB, Gothenburg, Sweden) were evaluated. The different composition of the evaluated bioinks is listed in [Table 1]. Printability tests were performed with a nozzle using a 3D printer (Inkredible; CELLINK AB, Gothenburg, Sweden) and a curvy bracelet model. Printing parameters were adjusted to give a smooth outflow of the bioink, and a total of five samples were printed for each composition. The samples were cross-linked with a 100 mM calcium chloride solution for 6 min. | Table 1: Composition of the three bioinks (Ink B23, Ink B42, and Ink B48) evaluated for printability using a nozzle. Beta tri-calcium phosphate, nanofibrillated cellulose/alginate (CELLINK)
Click here to view |
Nozzle design
Coaxial needles (Ramé-Hart Instruments Co., Succasunna, NJ, USA) [Figure 2]a using three different nozzle gauge combinations (22–18, 22–16, and 22–14) were tested for extrusion printing of the bioinks. [Figure 2]b illustrates a schematic of the nozzle configuration, and [Figure 2]c presents the measurements of the different gauge combinations tested. Distance A, i.e., the inner diameter of the coaxial needle, was maintained at 0.406 mm throughout all the printing procedures in the present study. Distance B, i.e., the outer diameter, was set to 0.129, 0.479, and 0.889 mm for the respective nozzle gauge combination tested. | Figure 2: (a) Image of coaxial needle (b) Description of nozzle configuration. Arrow A represents the inner diameter and arrow B represents the outer diameter of the respective nozzle (c) Table of combinations tested; remake according to the information provided by Ramé-Hart Instrument Co. (www.ramehart.com/pdf/needles.pdf)
Click here to view |
Cell culture preparation
The hAD-MSCs used for printing were RoosterVial hBM-10M (MSC-001) (RoosterBio Inc., Frederick, MD, USA) purchased in aliquots of 10 × 106 cells. Cell expansion was performed in accordance with the manufacturer's protocol. In detail, an expansion medium was prepared by mixing RoosterBio media booster and basal medium, respectively. The cell aliquot was brought to 37°C and transferred to a 10 ml centrifuge tube. Four milliliters of expansion medium was added drop-wise to the cell suspension. The solution was then centrifuged at 200 g for 5 min. The supernatant was aspired, and the cell pellet re-suspended in 2 ml expansion medium and incubated for 5 days at 37°C and 5% CO2. The cell medium was changed at day 3. A concentration of 4 × 106 cells/ml ink yielded a total number of 148 constructs with 3.3 ml ink during one print process.
Three-dimensional printing
Printing without adipose stem cells
Grids were printed using the 22–16 gauge combination previously described and a combination of Ink B23 or B42 as an inner ink and CELLINK as an outer ink. The mineral ink was inserted to printer head 1 (PH 1) and the CELLINK was connected to printer head 2 (PH 2). Grids in two layers, 9 mm × 9 mm, were printed using a pressure of 105 kPa for PH 1 and 55 kPa for PH 2. Completed grids were cross-linked with CaCl2, 100 mM (Sigma-Aldrich, Stockholm, Sweden) for 6 min before being transferred into Hanks balanced salt solution (HBSS; GIBCO) and incubated at 37°C in 5% CO2 overnight. Following incubation, stability of the printed grids was evaluated by assessing the degree of maintained grid outline.
Bioprinting with adipose stem cells
Printing was done within 2 h after cell preparation, using the 3D printer (Inkredible; CELLINK AB, Gothenburg, Sweden). Before mixing the hAD-MSCs with Ink B23 or CELLINK, cell confluence was evaluated by confocal light microscopy. Confluent cells were washed with 6 ml Dulbecco's phosphate-buffered saline (GIBCO) and incubated with 6 ml TrypLE™ (GIBCO) at 37°C for 1 min. Cell de-attachment was the verified by confocal light microscopy, and TrypLE™ activity was quenched by adding 6 ml spent expansion media. For the printing, the mineral ink was inserted to PH 1 and the bioink containing cells to PH 2. Grids in two layers, 6 mm × 6 mm, were printed using the combinations of inks and bioinks previously evaluated. A pressure of 105 kPa (PH 1) and 55 kPa (PH 2) was then applied. For the evaluation of printing pressure effects on the cells in the Ink B23, two different pressures on PH 2 were tested, 75 kPa and 85 kPa, respectively. Completed grids were cross-linked with CaCl2, 100 mM (Sigma-Aldrich, Stockholm, Sweden) for 6 min before being transferred into expansion medium and incubated at 37°C in 5% CO2.
Viability control
Cell viability was measured 24 h after printing using a Olympus Microscope with ×10 objective. The cells were stained using a Live/Dead Cell Imaging Kit (Invitrogen), a two-component imaging kit which stains viable cells with calcein-AM and dead cells with BOBO-3 Iodine. The staining was performed in accordance with the manufacturer's protocol. In brief, the printed constructs were washed with HBSS for 30 min at 37°C. The two different components in the live/dead kit were mixed together to make a 2 × working solution. The solution was mixed with an equal volume of HBSS and added to cover the printed constructs. The constructs were incubated for 1 h at 37°C. The staining solution was removed and washed with HBSS for 1 h at 37°C before imaging. Viable cells were imaged by exciting at 488 nm and the dead cells at 570 nm. Images were processed using the image processing software Image J (NIH, LOCI; University of Wisconsin, WI, USA).
Results | |  |
Inks and bioinks
Among the three evaluated combinations of bioinks (Ink B23, Ink B42, and Ink B48), two combinations of bioinks resulted in good printability for extrusion bioprinting, Ink B23 (23% ground β-TCP; 77% CELLINK) as well as Ink B42 (42% ground β-TCP; 58% CELLINK) [Figure 3]a. The remaining combination, Ink B48 (48% ground β-TCP; 52% CELLINK) [Figure 3]b, caused clogging and negative pressure effects on the prints; hence, this combination was excluded from further tests in this study. Ink B23 and Ink B42 were both considered suitable for further testing. | Figure 3: Printability test of (a) Ink B42 and (b) Ink B48 using three-dimensional printer (Inkredible) and curvy bracelet model for the two bioink compositions
Click here to view |
Nozzle design
The suitable bioinks from the printability tests, i.e., Ink B23 and Ink B42, were further evaluated using a coaxial needle. Out of the three nozzle configurations available only two were successful in printing an adequate amount of β-TCP + CELLINK. Printing with the 22-18 gauge combination was unsuccessful due to the narrow path for the outer ink (B distance) [Figure 2]b, which means that a much higher pressure would be needed for printing than is currently allowed by the 3D printer (Inkredible). The remaining gauge combinations, 22–16 and 22–14, both proved to be good for printing. However, due to the smaller total opening, the 22–16 gauge combination was shown to give a better printing resolution. Furthermore, when the 22–16 and the 22–14 gauge coaxial needles were compared with regard to printing of β-TCP + CELLINK, the former revealed a better stability and was used for further experiments.
Bioprinting
Stability tests were performed by printing grids with the 22–16 gauge nozzle combination. This was carried out in two different settings, either by using the two bioinks previously determined best for printing, namely Ink B23 or Ink B42 and CELLINK [Figure 4]a and [Figure 4]d or by using the two bioinks without any additional CELLINK [Figure 4]g. Intact grids were possible to print using the Ink B23 or Ink B42 as the inner ink and CELLINK as the outer ink [Figure 4]b and [Figure 4]e. Stable grids were also possible to print without using CELLINK, with Ink B42 as the inner ink and Ink B23 as the outer ink [Figure 4]h. All these combinations were shown to be printable with a maintained grid outline and stability after one night in HBSS [Figure 4]c, [Figure 4]f, and [Figure 4]i. Hence, all combinations were used for further cell viability experiments. | Figure 4: Printing process, completed grids, and stability test after one night in Hanks balanced salt solution, using either Ink B23 or Ink B42 as inner ink and CELLINK as outer ink (a-f) or by using the two bioinks, Ink B42 and Ink B23 as the inner and outer ink, respectively (g-i)
Click here to view |
Viability control
hAD-MSCs were mixed together with CELLINK and printed as an outer ink with either the Ink B23 or the Ink B42 as an inner ink. hAD-MSCs were also mixed together with the Ink B23 and printed as an outer ink (at two different printing pressures) using B42 as an inner ink. Cell viability was evaluated after 24 h. The cell viability test following printing revealed a higher viability in the grids enclosing Ink B23 as the inner ink and CELLINK + hAD-MSCs as the outer ink as compared to the grids with B42 as the inner ink and CELLINK + hAD-MSCs as the outer ink. Nearly 80% of the cells were viable in the grids comprising Ink B23 whereas the corresponding figure was 65% for the grids with Ink B42 [Figure 5]a. Grids comprising Ink B42 as an inner ink and hAD-MSCs mixed with Ink B23 as an outer ink had a higher cell viability when printed at 75 kPa compared to 85 kPa [Figure 5]b. The grids printed with lower pressure retained a cell viability of about 65% which was similar to the grids enclosing Ink B42 as an inner ink and CELLINK and hAD-MSCs as an outer ink. Higher pressures for printing the cells resulted in lower viability, with a viability of <60%. | Figure 5: Cell viability test of printed grids comprising human adipose tissue derived mesenchymal stem cells. (a) Ink B23 or Ink B42 as the inner ink and CELLINK + human adipose tissue-derived mesenchymal stem cells as the outer ink. (b) Ink B42 as the inner ink and CELLINK + human adipose tissue-derived mesenchymal stem cells as the outer ink, at 75 kPa and 85 kPa, respectively
Click here to view |
Discussion | |  |
The primary aim of the present study was to identify and develop a bioprinting process with a predictable outcome, producing a stable scaffold with viable cells. These factors are important when introducing bioprinting into the clinical setting, such as to use for reconstructing major skeletal defects in the head-and-neck area. The craniomaxillofacial skeleton is load bearing, implying that the printed construct must be stable. While hydrogels are commonly used materials in bioprinting, they are not inherently load bearing. Cross-linking the hydrogel is a common procedure to increase its stability. However, cross-linking is not sufficient to fulfill the needs for a skeletal construct. In the present study, a synthetic material, β-TCP, was chosen as the scaffold to help achieve a stable printed construct. β-TCP is today used as a bone substitute to restore minor and intermediate-sized defects in the maxillofacial area with encouraging results.[24],[25],[26] However, it must be noted that, in larger defects, β-TCP without cells is not sufficient for predictable bone healing.[26] Therefore, this study evaluated three different compositions of β-TCP and CELLINK and found that bioinks, comprising β-TCP of 23% and 42%, respectively, were printable with a stable structure using an extrusion 3D printer. Another issue of importance is the printing process itself. When a construct with the purpose to both deliver cells and acting as a scaffold to bridge a defect, both the cells and the scaffold must be printed concurrently, and their cellular viability must be maintained. The type of bioprinter may be essential; however, the 3D printer used in this study fulfilled the requirements and produced constructs at both 75 kPa and 85 kPa.
Earlier studies using mono-needles revealed substantial limitations, especially regarding the time required for printing.[35] Thus, to overcome these problems, the authors used a coaxial needle, making it possible to print both the cells as well as the scaffold simultaneously. Evaluation of different nozzle gauge combinations of the coaxial needle revealed that a coaxial needle with an inner diameter of 0.406 mm and an outer diameter of 0.479 mm, especially the 22–16 nozzle configuration, was most suitable for the purpose of our study. This as the 22–18 nozzle configuration resulted in a poorer printing resolution where the structure of the grid was not visible and the structural properties were insufficient. In addition, the cell laden layer was lost during movement of the grid. Furthermore, although the 22–14 gauge nozzle combination proved to be suitable for printing, it resulted in decreased stability of the printed grids when compared to the 22–16 coaxial needle. Hence, the authors decided to use the 22–16 gauge nozzle combination for subsequent experiments of this study.
Finally, the problem to maintain cell viability of the printed grids was studied. Our results indicate that the number of human adipose stem cells in the preparation was sufficient, which has been concluded by other studies.[26] However, the results with an increased number of cells to determine their optimal number in the print have yet to be investigated. Another finding was that the printing pressure was of importance. A higher pressure resulted in an increased number of dead cells after 24 h. In addition, the viscosity of the bioink used for printing the scaffold seemed to be important since the ratio of living versus dead cells decreased with higher viscosity of the scaffold ink. However, the effect of the dead versus living cells needs to be further explored since cell degradation may release intracellular products that may both stimulate and hamper the activity of the remaining living cells. In the present series of experiments, the size of the printed grids was small. However, cell viability in larger grids is of importance to further investigate as these may be used to replace complex-shaped bony defects in the facial skeleton. Combinations using other printing materials or processes such as a sacrificial matrix material which is intended to stabilize the larger construct during a predefined period may be a possibility. Different printing pressures and printing speed are other important aspects to consider for further research. Taken together, although the results in this study look promising, the authors want to further develop the technique for better dimensional stability, cell viability, and reproducibility so it can be put to use in vivo and then in clinical practise.
Conclusions | |  |
This study indicates that using a coaxial needle enables printing of a stable scaffold with viable cells. Furthermore, cell viability is maintained after the bioprinting process.
Acknowledgments
The authors would like to acknowledge Dr. Michaela Schettino and Dr. Hector Martinez for their assistance in conducting the experiments in this study.
Financial support and sponsorship
The Local Research and Development Council Region Västra Götaland, The Knut and Alice Wallenberg Foundation, is kindly acknowledged for financial support in the framework of the Wallenberg Wood Science Center.
Conflicts of interest
There are no conflicts of interest.
References | |  |
1. | Bianchi B, Ferri A, Ferrari S, Leporati M, Copelli C, Ferri T, et al. Mandibular resection and reconstruction in the management of extensive ameloblastoma. J Oral Maxillofac Surg 2013;71:528-37. |
2. | Pogrel MA, Podlesh S, Anthony JP, Alexander J. A comparison of vascularized and nonvascularized bone grafts for reconstruction of mandibular continuity defects. J Oral Maxillofac Surg 1997;55:1200-6. |
3. | Goh BT, Lee S, Tideman H, Stoelinga PJ. Mandibular reconstruction in adults: A review. Int J Oral Maxillofac Surg 2008;37:597-605. |
4. | Dasmah A, Thor A, Ekestubbe A, Sennerby L, Rasmusson L. Particulate vs. block bone grafts: three-dimensional changes in graft volume after reconstruction of the atrophic maxilla, a 2-year radiographic follow-up. J Craniomaxillofac Surg 2012;40:654-9. |
5. | Schliephake H. Revascularized tissue transfer for the repair of complex midfacial defects in oncologic patients. J Oral Maxillofac Surg 2000;58:1212-8. |
6. | Hidalgo DA. Fibula free flap: A new method of mandible reconstruction. Plast Reconstr Surg 1989;84:71-9. |
7. | Thankappan K, Trivedi NP, Subash P, Pullara SK, Peter S, Kuriakose MA, et al. Three-dimensional computed tomography-based contouring of a free fibula bone graft for mandibular reconstruction. J Oral Maxillofac Surg 2008;66:2185-92. |
8. | Sàndor GK, Nish IA, Carmichael RP. Comparison of conventional surgery with motorized trephine in bone harvest from the anterior iliac crest. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003;95:150-5. |
9. | Ling XF, Peng X. What is the price to pay for a free fibula flap? A systematic review of donor-site morbidity following free fibula flap surgery. Plast Reconstr Surg 2012;129:657-74. |
10. | Sieg P, Taner C, Hakim SG, Jacobsen HC. Long-term evaluation of donor site morbidity after free fibula transfer. Br J Oral Maxillofac Surg 2010;48:267-70. |
11. | Nathan SS, Athanasian E, Boland PJ, Healey JH. Valgus ankle deformity after vascularized fibular reconstruction for oncologic disease. Ann Surg Oncol 2009;16:1938-45. |
12. | Gimbel M, Ashley RK, Sisodia M, Gabbay JS, Wasson KL, Heller J, et al. Repair of alveolar cleft defects: Reduced morbidity with bone marrow stem cells in a resorbable matrix. J Craniofac Surg 2007;18:895-901. |
13. | Löfstrand J, Nyberg M, Karlsson T, Thórarinsson A, Kjeller G, Lidén M, et al. Quality of life after free fibula flap reconstruction of segmental mandibular defects. J Reconstr Microsurg 2018;34:108-20. |
14. | Langer R, Vacanti JP. Tissue engineering. Science 1993;260:920-6. |
15. | Sándor GK, Suuronen R. Combining adipose-derived stem cells, resorbable scaffolds and growth factors: An overview of tissue engineering. J Can Dent Assoc 2008;74:167-70. |
16. | Thesleff T, Lehtimäki K, Niskakangas T, Mannerström B, Miettinen S, Suuronen R, et al. Cranioplasty with adipose-derived stem cells and biomaterial: A novel method for cranial reconstruction. Neurosurgery 2011;68:1535-40. |
17. | Gimble JM, Zvonic S, Floyd ZE, Kassem M, Nuttall ME. Playing with bone and fat. J Cell Biochem 2006;98:251-66. |
18. | Chamberlain G, Fox J, Ashton B, Middleton J. Concise review: Mesenchymal stem cells: their phenotype, differentiation capacity, immunological features, and potential for homing. Stem Cells 2007;25:2739-49. |
19. | Song L, Tuan RS. Transdifferentiation potential of human mesenchymal stem cells derived from bone marrow. FASEB J 2004;18:980-2. |
20. | Pittenger MF, Mackay AM, Beck SC, Jaiswal RK, Douglas R, Mosca JD, et al. Multilineage potential of adult human mesenchymal stem cells. Science 1999;284:143-7. |
21. | Wolf NS, Penn PE, Rao D, McKee MD. Intraclonal plasticity for bone, smooth muscle, and adipocyte lineages in bone marrow stroma fibroblastoid cells. Exp Cell Res 2003;290:346-57. |
22. | Zimmermann CE, Gierloff M, Hedderich J, Açil Y, Wiltfang J, Terheyden H. Survival of transplanted rat bone marrow-derived osteogenic stem cells in vivo. Tissue Eng Part A 2011;17:1147-56. |
23. | Lindroos B, Suuronen R, Miettinen S. The potential of adipose stem cells in regenerative medicine. Stem Cell Rev Rep 2011;7:269-91. |
24. | Mesimäki K, Lindroos B, Törnwall J, Mauno J, Lindqvist C, Kontio R, et al. Novel maxillary reconstruction with ectopic bone formation by GMP adipose stem cells. Int J Oral Maxillofac Surg 2009;38:201-9. |
25. | Sándor GK, Tuovinen VJ, Wolff J, Patrikoski M, Jokinen J, Nieminen E, et al. Adipose stem cell tissue-engineered construct used to treat large anterior mandibular defect: A case report and review of the clinical application of good manufacturing practice-level adipose stem cells for bone regeneration. J Oral Maxillofac Surg 2013;71:938-50. |
26. | Sándor GK, Numminen J, Wolff J, Thesleff T, Miettinen A, Tuovinen VJ, et al. Adipose stem cells used to reconstruct 13 cases with cranio-maxillofacial hard-tissue defects. Stem Cells Transl Med 2014;3:530-40. |
27. | Leong KF, Cheah CM, Chua CK. Solid freeform fabrication of three-dimensional scaffolds for engineering replacement tissues and organs. Biomaterials 2003;24:2363-78. |
28. | Schliephake H, Zghoul N, Jäger V, van Griensven M, Zeichen J, Gelinsky M, et al. Effect of seeding technique and scaffold material on bone formation in tissue-engineered constructs. J Biomed Mater Res A 2009;90:429-37. |
29. | Haimi S, Moimas L, Pirhonen E, Lindroos B, Huhtala H, Räty S, et al. Calcium phosphate surface treatment of bioactive glass causes a delay in early osteogenic differentiation of adipose stem cells. J Biomed Mater Res A 2009;91:540-7. |
30. | Hämmerle CH, Olah AJ, Schmid J, Flückiger L, Gogolewski S, Winkler JR, et al. The biological effect of natural bone mineral on bone neoformation on the rabbit skull. Clin Oral Implants Res 1997;8:198-207. |
31. | Kokemueller H, Spalthoff S, Nolff M, Tavassol F, Essig H, Stuehmer C, et al. Prefabrication of vascularized bioartificial bone grafts in vivo for segmental mandibular reconstruction: experimental pilot study in sheep and first clinical application. Int J Oral Maxillofac Surg 2010;39:379-87. |
32. | Arias-Gallo J, Maremonti P, González-Otero T, Gómez-García E, Burgueño-García M, Chamorro Pons M, et al. Long term results of reconstruction plates in lateral mandibular defects. Revision of nine cases. Auris Nasus Larynx 2004;31:57-63. |
33. | Wolff J, Sándor GK, Miettinen A, Tuovinen VJ, Mannerström B, Patrikoski M, et al. GMP-level adipose stem cells combined with computer-aided manufacturing to reconstruct mandibular ameloblastoma resection defects: Experience with three cases. Ann Maxillofac Surg 2013;3:114-25.  [ PUBMED] [Full text] |
34. | Axpe E, Oyen ML. Applications of Alginate-Based Bioinks in 3D Bioprinting. Int J Mol Sci 2016;17:1-11. |
35. | Tomasina C, Bodet T, Mota C, Moroni L, Camarero-Espinosa S. Bioprinting Vasculature: Materials, Cells and Emergent Techniques. Materials (Basel) 2019;12:1-42. |
[Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
[Table 1]
|